After Sounding Rocket flight (US36.181.) on 18 August1999 the SEM was prepared for calibration at NIST. The SEM had been stored in a dry Nitrogen sealed bag after flight, and was prepared for transport by pumping in an Ion Pump chamber for one week prior to shipment, and then sealed in dry nitrogen for transport to NIST.
Thursday October 21 1999
The SEM instrument was mounted on the carrier for Beam Line 9 (BL9) in the horizontal position. The top cover of the SEM was removed as was the aperture assembly and the front assembly.
The filter was orriented with the serial# in the lower RH corner and the grating serial# in the upper LH corner as viewed from the front of the SEM.
The wiring harnesses were connected and a ground strap was placed between one of the feedthrough BNCs to the chassis of the SEM. Shorting the BNC to the structure case reduced the noise from about 7cps to about 3cps.
The pumps were started to pump down the system
Friday October 22 1999
The data collection system for BL9 has changed from a microVax to a Macintosh G3. Work on the interaction between CALROB2.BAS (on our PC) and the Labwindows system on the Mac produce and interactive system that displays the QE in real time.
Sunday October 24
Write NISTSCAN.BAS for the PC to provide autonomous, highresolution wavelength scans on BL9 in conjunction with the MacG3 controller.
Monday October 25
Chamber pressure is at 5¥10-8 T
| X | Q | Y | F |
| 0.4635 | 0.9905 | 0.9415 | 0.5000 |
10:50 Start Calibration: T=148: Grating 0°
Calibration data sets:
1A
1B
1C (wavelength scan problem in 1C)
12:10
Scan data sets
2A (340 170 Å, step 1Å)
2B (340 170 Å, step 3Å1
14:30 New Beam
3C (340 170Å, step 3Å)
Calibration data set
4A (304Å point added)
note: We still get turn-up at small wavelengths in the calibration runs, not as much in the scans. Rob modified the scan program on the mac to measure the monitor dark for every scan position.
Scan data set
5A (340 160 Å step 3Å) (Monitor darks)
6A (340 160 Å step 5Å) (Record fluxes)
note: We seem to be seriously photon flux / statistics limited shortward of 200Å, so for calibrations it is probably a good idea to increase nsample (currentlt 20).
Try with an e-beam energy of 1331MeV (was 180 MeV)
Scan data set
7A (340 160Å step 5Å) (Beam energy 331 MeV)
Tuesday 26 October
9:49 Beam energy 180 MeV
Scan data set
1A (340 170 Å step 5Å)
Calibration data set
2A (BNC ground strap on)
2B (BNC ground strap off)
Start doing grating scans: 0.1° corresponds to a change in micrometer position of 0.0144".
| X | Q | F | Y |
| 0.41785 | 1.0049 | 0.9415 | 0.5000 |
Scan data set
3A (340 175 Å step 5Å: Q = + 0.1 °)
| X | Q | F | Y |
| 0.3826 | 1.0193 | 0.9415 | 0.5000 |
Scan data set
3B (340 175 Å step 5Å: Q = + 0.2 °)
| X | Q | Y | Y |
| 0.35114 | 1.0337 | 0.9415 | 0.5000 |
Scan data set
3C (340 175 Å step 5Å: Q = + 0.3 °)
13:15 New Beam
| X | Q | F | Y |
| 0.49537 | 0.9761 | 0.9415 | 0.5000 |
Scan data set
3D (340 175 Å step 5Å: Q = - 0.1 °)
| X | Q | F | Y |
| 0.53325 | 0.9617 | 0.9415 | 0.5000 |
Scan data set
3E (340 175 Å step 5Å: Q = - 0.2 °)
| X | Q | F | Y |
| 0.57133 | 0.9473 | 0.9415 | 0.5000 |
Scan data set
3F (340 175 Å step 5Å: Q = - 0.3 °)
note: connect a BNC T into the channel 2 data line, and use a frequency counter for alignment. This saves having to move the GSE box every time to the BL. We checked that it has no effect on the recorded counts or noise.
| X | Q | F | Y |
| 0.4635 | 0.9905 | 0.9415 | 0.5000 |
Calibration Scan data set
4A (340 175 Å step 5Å: Q = 0.0 °)
Calibration data set
4B
Calibration Scan data set
5A (360 260 Å step 5Å)
Scan data set
5B (360 260Å step 5Å)
Calibration data set
6A (repeat 5 measurements at 290 Å)
Scan data set
6B (repeat 5 measurements at 290 Å)
note: The discrepancy between the displayed scan and calibration values is that when I send the data from the PC to the mac I do not subtract the dark counts. Modify the program to send dark subtracted counts to the Mac.
Calibration data set
7A (repeat 5 measurements at 290 Å)
This now gives good agreement wirh 6B
Note: the problem does not effect the data recorded on the PC, ONLY the display on the Mac
Wednesday 27 October
Scan X micrometer to determine beam width. Center was at 0.46350. Use frequency counter to make measurements and alignments.
|
|
|
| 0.250 | 25 |
| 0.30 | 38 |
| 0.325 | 277 |
| 0.350 | 812 |
| 0.375 | 1438 |
| 0.400 | 2100 |
| 0.425 | 2689 |
| 0.450 | 3059 |
| 0.475 | 2900 |
| 0.500 | 2410 |
| 0.525 | 1755 |
| 0.550 | 1147 |
| 0.575 | 505 |
| 0.600 | 155 |
| 0.625 | 38 |
| 0.650 | 27 |
| 0.70 | 26 |
| dark | 25 |
| FWHM | 4.034mm |
| Base | 7.1mm |
| Beam Size | 3.1mm |
| X | Q | F | Y |
| 0.43554 | 0.9970 | 0.9415 | 0.5000 |
Scan data set
1A (340 175 Å step 5Å: Q = + 0.05 °)
| X | Q | F | Y |
| 0.38060 | 1.0193 | 0.9415 | 0.5000 |
Scan data set
2B(340 175 Å step 5Å: Q = + 0.2 °)
Break Vacuum and rotate the SEM 90° CCW
Rotate the F micrometer 90° to make it easier to read.
| X | Q | F | Y |
| 0.5000 | 0.9587 | 0.4973 | 0.9671 |
Calibration data set
3A (dark counts high because the SEM does not shield the Ion pump)
3B
Scan data set
4A (340 175 Å step 5Å: F= + 0.0 °)
| X | Q | F | Y |
| 0.5000 | 0.9587 | 0.57183 | 0.9959 |
Scan data set
4B (340 175 Å step 5Å: F= + 0.2 °)
| X | Q | F | Y |
| 0.5000 | 0.9587 | 0.42114 | 0.9383 |
Scan data set
4B (340 175 Å step 5Å: F= - 0.2 °)
Thursday 28 October
| X | Q | F | Y |
| 0.5000 | 0.9587 | 0.53647 | 0.9815 |
Scan data set
1A (340 175 Å step 5Å: F= + 0.1 °)
| X | Q | F | Y |
| 0.5000 | 0.9587 | 0.46001 | 0.9527 |
Scan data set
1B (340 175 Å step 5Å: F= - 0.1 °)
| X | Q | F | Y |
| 0.5000 | 0.9587 | 0.49769 | 0.9671 |
Calibration data set
2A (340 175 Å step 5Å: F= 0.0 °)
Break Vacuum
Rotate Grating; serial # is now in lower left viewed from front.
New Beam
| X | Q | F | Y |
| 0.5000 | 0.9573 | 0.9652 | 0.4946 |
Calibration data set
3A (340 175 Å step 5Å: F= 0.0 °)
note: Channel 3 efficiency is back!
Scan data set
3A (340 175 Å step 5Å: F= 0.0 °)
| X | Q | F | Y |
| 0.5000 | 0.9573 | 0.9940 | 0.57128 |
Scan data set
4A (340 175 Å step 5Å: F= +0.2 °)
| X | Q | F | Y |
| 0.5000 | 0.9573 | 0.9573 | 0.57128 |
Scan data set
4B (340 175 Å step 5Å: F= +0.2 °) (check alignment)
| X | Q | F | Y |
| 0.5000 | 0.9573 | 0.9364 | 0.42097 |
Scan data set
4B (340 175 Å step 5Å: F= -0.2 °)
Beam Geometry measurement Y-Scan
|
|
|
| 0.350 | 4 |
| 0.375 | 8 |
| 0.400 | 47 |
| 0.425 | 88 |
| 0.450 | 128 |
| 0.475 | 153 |
| 0.500 | 153 |
| 0.525 | 151 |
| 0.550 | 115 |
| 0.575 | 72 |
| 0.600 | 33 |
| 0.625 | 6 |
| 0.650 | 5 |
| dark | 4 |
| FWHM | 3.854 mm |
| Base | 6.167 mm |
| Beam Size | 2.313 mm |
Friday 29 October
Break vacuum and rotate SEM to horizontal position
Reinstall in BL9 and pump
Friday AM. Is used for maintainance of the ring.
| X | Q | F | Y |
| 0.4742 | 0.98115 | 0.9575 | 0.5000 |
Calibration data set
1A (340 175 Å step 5Å: Q = 0.0 °)
Scan data set
2A (340 175 Å step 5Å: Q = 0.0 °)
| X | Q | F | Y |
| 0.42240 | 1.0100 | 0.9575 | 0.5000 |
Scan data set
2B (340 175 Å step 5Å: Q = +0.2 °)
| X | Q | F | Y |
| 0.54748 | 0.9524 | 0.9575 | 0.5000 |
Scan data set
3C (340 175 Å step 5Å: Q = -0.2 °)
Beam Geometry measurement X-Scan
|
|
|
| 0.275 | 2 |
| 0.300 | 4 |
| 0.325 | 18 |
| 0.350 | 47 |
| 0.375 | 86 |
| 0.400 | 123 |
| 0.425 | 146 |
| 0.450 | 151 |
| 0.475 | 138 |
| 0.500 | 109 |
| 0.525 | 71 |
| 0.550 | 34 |
| 0.575 | 11 |
| 0.600 | 3 |
| 0.625 | 2 |
| dark | 1.5 |
| FWHM | 3.917 mm |
| Base | 6.438 mm |
| Beam Size | 2.521 mm |
Break Vacuum and rotate SEM 90° CW
Pump
Monday November 1 1999.
| X | Q | F | Y |
| 0.500 | 0.9791 | 0.9740 | 0.5022 |
Calibration data set
1A (F = 0.0 °)
Scan data set
2A (340 170 Å step 5Å: F = 0.0 °)
| X | Q | F | Y |
| 0.500 | 0.9791 | 1.0028 | 0.57720 |
Scan data set
2B (340 170 Å step 5Å: F = +0.2 °)
| X | Q | F | Y |
| 0.500 | 0.9791 | 0.9452 | 0.42574 |
Scan data set
2B (340 170 Å step 5Å: F = -0.2 °)
Beam Geometry measurement Y-Scan
|
|
|
304Å |
| 0.350 | 2 | 2 |
| 0.375 | 2 | 2 |
| 0.400 | 30 | 10 |
| 0.425 | 67 | 34 |
| 0.450 | 103 | 60 |
| 0.475 | 132 | 85 |
| 0.500 | 135 | 95 |
| 0.525 | 134 | 95 |
| 0.550 | 109 | 88 |
| 0.575 | 71 | 62 |
| 0.600 | 35 | 38 |
| 0.625 | 5 | 12 |
| 0.650 | 2 | 2 |
| 0.675 | 2 | |
| dark | 2 | 2 |
| Grating | 3.867 mm | 3.812 mm |
| Beam Size | 2.309 mm | 2.313 mm |
Break Vacuum
Remove grating/filter assembly
Remove Solar wind deflector assembly
| X | Q | F | Y |
| 0.500 | 0.9791 | 0.9740 | 0.5022 |
Calibration data set
1A (F = 0.0 ° Channel 2 only no grating)
Beam Geometry measurement Y-Scan, 304 Å
|
|
|
| 0.325 | 10 |
| 0.350 | 11 |
| 0.375 | 147 |
| 0.400 | 380 |
| 0.425 | 620 |
| 0.450 | 863 |
| 0.475 | 997 |
| 0.500 | 1016 |
| 0.525 | 1030 |
| 0.550 | 1042 |
| 0.575 | 1030 |
| 0.600 | 793 |
| 0.625 | 540 |
| 0.650 | 285 |
| 0.675 | 46 |
| 0.700 | 13 |
| 0.725 | 12 |
| dark | 7 |
| FWHM | 5.40mm |
| Base | 8.075 mm |
| Beam Size | 2.675 mm |
Break Vacuum
Rotate SEM to Horizontal
No grating, filter or Solar wind deflector assembly
| X | Q | F | Y |
| 0.500 | 0.98115 | 0.9575 | 0.5022 |
Beam Geometry measurement X-Scan, 304 Å
|
|
|
| 0.250 | 10 |
| 0.275 | 20 |
| 0.300 | 190 |
| 0.325 | 730 |
| 0.350 | 1580 |
| 0.375 | 2530 |
| 0.400 | 3400 |
| 0.425 | 4000 |
| 0.450 | 4200 |
| 0.475 | 4185 |
| 0.500 | 4040 |
| 0.525 | 3670 |
| 0.550 | 2970 |
| 0.575 | 2100 |
| 0.600 | 1230 |
| 0.625 | 535 |
| 0.650 | 136 |
| 0.675 | 38 |
| 0.700 | 28 |
| 0.725 | 23 |
| 0.750 | 18 |
| 0.775 | 15 |
| 0.800 | 12 |
| 0.825 | 10 |
| dark | 7 |
| FWHM | 5.334mm (diode) |
| Base | 8.367 mm |
| Beam Size | 3.033 mm |
Break Vacuum
Reinstall graing in ORIGINAL orrientation
Replace deflector assembly
SEM in Horizontal poition
Start pumps
Reset date/time on DAQ PC.
Tuesday November 2 1999.
| X | Q | F | Y |
| 0.46854 | 0.9797 | 0.95055 | 0.5000 |
Beam Geometry measurement X-Scan
|
|
|
| 0.300 | 1 |
| 0.325 | 7 |
| 0.350 | 27 |
| 0.375 | 55 |
| 0.400 | 83 |
| 0.425 | 110 |
| 0.450 | 132 |
| 0.475 | 135 |
| 0.500 | 117 |
| 0.525 | 90 |
| 0.550 | 62 |
| 0.575 | 34 |
| 0.600 | 13 |
| 0.625 | 3 |
| 0.650 | 2 |
| 0.675 | 1 |
| dark | 1 |
| FWHM | 3.975 mm (diode) |
| Base | 7.075 mm |
| Beam Size | 3.10 mm |
Calibration data set
1A (Q = 0.0 °)
Scan data set
2A (340 170 Å step 5Å: Q = 0.0 °)
Break Vacuum
Move SEM to vertical
New Beam
Note: It is Very windy outside, and the risk of a power surge forces us to dump the beam and call it a day.
Wednesday November 3 1999.
Note: We are trying to find the excess dark counts that give the low wavelength tails.
Change the wavelength file to look at only: sparse wavelengths:
330, 304, 290, 270, 215 & 175 Å
Run with No BEAM! Open BL valves set up exactly as normal but disconnect shutter
Calibration data set
1A (No beam, shutter closed, Ion pump open, GND strap on, sparse l set, Vertical (not alligned))
Note: The Kiethley 617 (diode monitor) does not give the same reading for signal / dark even though nothing changes physically. Run the 617 manually, it seems to be just statistical variation.
Calibration data set
1B (No beam, shutter open, Ion pump open, GND strap on, sparse l set, Vertical (not alligned))
1C (No beam, shutter changing, Ion pump open, GND strap on, sparse l set, Vertical (not alligned))
1D (same as 1C)
note: No systematic effects seen
Ion pump closed, noise goes down.
Calibration data set
2A (No beam, shutter changing, Ion pump closed, GND strap on, sparse l set, Vertical (not alligned))
note: no assymetry seen
Calibration data set
3A (No beam, shutter changing, Ion pump closed, GND strap off, sparse l set, Vertical (not alligned))
Inject beam
Rough Vertical alignment
Shutter normal
Remove external frequency counter ground
Calibration data set
4A (beam, shutter changing, Ion pump closed, GND strap off, sparse l set, Vertical, frequency counter disconnected)
4B (beam, shutter changing, Ion pump closed, GND strap off, sparse l set, Vertical, frequency counter connected)
4C (beam, shutter changing, Ion pump open, GND strap off, sparse l set, Vertical, frequency counter disconnected)
4D (exactly the same as 4C, but dont restart control program)
4E (beam, shutter changing, Ion pump open, GND strap on, sparse l set, Vertical, frequency counter disconnected transient as GND connected)
Note: no tail seen in any of the 4A 4E data sets!
Full SEM Vertical alignment
| X | Q | F | Y |
| 0.5000 | 0.9788 | 0.9764 | 0.51867 |
Calibration data set
5A (sparse l)
5B (full l)
Note: no tail seen in the 5A,B data sets!
Dismount SEM from BL9, return grating to high efficiency position. Pack in dry nitrogen and ship back to USC.
The linear size of the beam from the SURFiii facility at BL-9 was measured by SEM by linear scans of the SEM channel-2 (0-order) detector over the beam. The SEM filter and grating were left in the instrument for the measurements.
The response of the SEM to linear scans is roughly trapizoidal, as expected for scanning a finite size beam over an aperture (the SEM grating). The size of the base of the trapezoid is B + W, where B is the beam size, and W the aperture width.
SEM Data
|
inches counts mm counts dark 0.250 25 6.3500 0 0.300 38 7.6200 13 0.325 277 8.2550 252 0.350 812 8.8900 787 0.375 1438 9.5250 1413 0.400 2100 10.160 2075 0.425 2689 10.795 2664 0.450 3059 11.430 3034 0.475 2900 12.065 2875 0.500 2410 12.700 2385 0.525 1755 13.335 1730 0.550 1147 13.970 1122 0.575 505 14.605 480 0.600 155 15.240 130 0.625 38 15.875 13 0.650 27 16.510 2 0.700 26 17.780 1 |
![]() |